SORIN Heater-Cooler System 3T 120V~/60Hz recall
SORIN Heater-Cooler System 3T 120V~/60Hz REF 16-02-85 Description of product: The St¿ckert Heater-Cooler System is an independent (i.e. independent of…
Why was SORIN Heater-Cooler System 3T 120V~/60Hz recalled?
The firm has become aware that due to a chemical reaction, the disinfectant/water preservative used to disinfect the device to inhibit/limit the growth micro-organisms, may be rendered ineffective. In addition, the firm is releasing a design upgrade to reduce the risk of potential emission of aerosols from the 3T.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
16-02-80 Stockert Heater-Cooler System 3T, 230V, Serial #s 16S10027 thru 16S16874 16-02-81 Stockert Heater-Cooler System 3T, 240 v/60 Hz, Serial #s 16S10743 thru 16S11708 16-02-82 Stockert Heater-Cooler System 3T, 208 v/60 Hz, Serial #s 16S10772 thru 16S16840 16-02-83 Stockert Heater-Cooler System 3T, 127V/60Hz, Serial #s 16S11455 thru 16S16509 16-02-85 Stockert Heater-Cooler System 3T, 120 v/60 Hz, Serial #s 16S10958 thru 16S16847 16-02-95 Stockert Heater-Cooler System 3T, 200V/50Hz/60Hz, Serial #s 16S12004 thru 16S16818
- Product
- SORIN Heater-Cooler System 3T 120V~/60Hz REF 16-02-85 Description of product: The St¿ckert Heater-Cooler System is an independent (i.e. independent of…
- Recalling firm
- LivaNova USA Inc.
- Quantity
- 8671
- Distribution
- Worldwide distribution - US Nationwide distribution including sate of PR and countries of Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Georgia, Germany, Greece, Guatemala, Guyana, Hong Kong, Hungary, India, Indonesia, Iran (Islamic Republic of), Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea (the Republic of), Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia (the former Yugoslav Republic of), Malaysia, Maldives, Malta, Mauritius, Mexico, Monaco, Mongolia, Montenegro, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines (the), Poland, Portugal, Qatar, Romania, Russian Federation (the), Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan (Province of China), Thailand, Trinidad and Tobago, Tunisia, Turkmenistan, Turkey, Ukraine, United Arab Emirates (the), United Kingdom, Uruguay, Uzbekistan, Vietnam.
- Recall date
- November 6, 2019
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)