Baxter Operating table TruSystem 7000 U recall
Baxter Operating table TruSystem 7000 U (MB), Product codes: a) REF 1604786; b) REF 1604788; c) REF 1723633
Why was Baxter Operating table TruSystem 7000 U recalled?
The firm received customer reports stating that the batteries and their connectors experienced electrical short-circuits and/or emitted smoke.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
a) REF 1604786, UDI/DI 00887761967533; b) REF 1604788, UDI/DI 00887761967526; c) REF 1723633, UDI/DI 00887761967519 ALL SERIAL NUMBERS
- Product
- Baxter Operating table TruSystem 7000 U (MB), Product codes: a) REF 1604786; b) REF 1604788; c) REF 1723633
- Recalling firm
- Baxter Healthcare Corporation
- Quantity
- 6045 units
- Distribution
- Worldwide distribution.
- Recall date
- October 30, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)