Phasitron5 Breathing Circuit recall
Phasitron5 Breathing Circuit (P5-10, P5-HC, P5-UC-10). component of Intrapulmonary Percussive Ventilation, used to provide airway clearance therapy
Why was Phasitron5 Breathing Circuit recalled?
As a result of product complaints, breathing circuits are experiencing discolored fluid which are formed when therapy was being given with specific medication through the nebulizer cup. Discoloration is caused by a reaction between medicated mist and an inadvertent/unapproved addition of a nickel coating on the spring component.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model Numbers: P5-10 P5-HC P5-UC-10 UDI-DI codes: Lot Numbers: 240321 240329 240412 240430 240509 240516 240614 240625 240627
- Product
- Phasitron5 Breathing Circuit (P5-10, P5-HC, P5-UC-10). component of Intrapulmonary Percussive Ventilation, used to provide airway clearance therapy
- Recalling firm
- Percussionaire Corporation
- Quantity
- 43,830 devices
- Distribution
- U.S.: PA, AR, MI, CA, TX, OH, NY, FL, IL, NE, OR, GA, WI, MO, OK, AL, WV, MT, CO, LA, VA, CT, SC, NH, NV, NC, NJ, MI, UT, ID, ME, MN, SD, ID, ND, KS, KY, TN, MD, VT, MA, RI, O.U.S.: Canada, Switzerland (all products shipped OUS remained in Sentec AG's (distributor, parent company of Percussionaire) control, were quarantined, and did not leave the warehouse).
- Recall date
- November 6, 2024
- Classification
- Class I
- Status
- Ongoing
- Agency
- FDA (device)