FDAClass IIOctober 30, 2024

Diasol Acid Concentrate recall: Labeling Error

Diasol Acid Concentrate REF G100325-10Dex 100

Why was Diasol Acid Concentrate recalled?

Due to labeling issue. The box label is different that the gallon label in that gallon label shows a lower Calcium concentration than the box label.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Catalog Number: G100325-10Dex100 UDI code: B467100325101004 Lot Number:PHS06011/061722

Product
Diasol Acid Concentrate REF G100325-10Dex 100
Recalling firm
Diasol, Inc
Quantity
852 gallon bottles
Distribution
U.S. Nationwide distribution in the state of AZ.
Recall date
October 30, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)