Diasol Acid Concentrate recall: Labeling Error
Diasol Acid Concentrate REF G100325-10Dex 100
Why was Diasol Acid Concentrate recalled?
Due to labeling issue. The box label is different that the gallon label in that gallon label shows a lower Calcium concentration than the box label.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Catalog Number: G100325-10Dex100 UDI code: B467100325101004 Lot Number:PHS06011/061722
- Product
- Diasol Acid Concentrate REF G100325-10Dex 100
- Recalling firm
- Diasol, Inc
- Quantity
- 852 gallon bottles
- Distribution
- U.S. Nationwide distribution in the state of AZ.
- Recall date
- October 30, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)