FDAClass IINovember 8, 2023

RayCare 5B and 6A including service packs- An oncology recall

RayCare 5B and 6A including service packs- An oncology information system used to support workflows and scheduling, as well as clinical information…

Why was RayCare 5B and 6A including service packs- An oncology recalled?

An issue was found in RayCare 5B, 6A, including service packs, where values will not be visible for documents in a read-only state when opened outside the Documents workspace when using document fields configured with single- or multi-selectable values.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI: (1)RayCare 5B SP1 (5.1.1.60246) 0735000201053220220316; (2)RayCare 5B SP3 (5.1.3.60023) 0735000201069320221027; (3) RayCare 6A (6.0.0.60553) 0735000201056320220617

Product
RayCare 5B and 6A including service packs- An oncology information system used to support workflows and scheduling, as well as clinical information…
Recalling firm
RAYSEARCH LABORATORIES AB
Quantity
2 sytems
Distribution
MD, TN
Recall date
November 8, 2023
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls