Why was Plum A+ Infusion Pump recalled?
Hospira has received reports of distal pressure sensor calibration drift on Plum A+ Infusers. The following error codes may be displayed during setup or infusion signifying possible calibration drift: E180/N180, E181/N181, E186/N186, E187/N187 or E346. Additionally, if the pressure sensor calibration has drifted, users may experience false distal occlusion alarms or distal occlusions that may not
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
List No. 11973; Serial Numbers: 12570148 , 12570150 , 12570154 , 12570163 , 12570170 , 12570177 , 12570204 , 12570211 , 12570212 , 12570255 , 12570270 , 12570319 , 12570359 , 12570385 , 12570413 , 12570425 , 12570455 , 12570487 , 12570662 , 12570695 , 12570854 , 12570915 , 12571080 , 12571243 , 12571437 , 12571631 , 12571692 , 12571776 , 12571826 , 12571835 , 12571837 , 12571842 , 12571877 , 12571880 , 12571886 , 12572317 , 12572369 , 12572586 , 12572909 , 12573547 , 12573567 , 12573837 , 12573990 , 12574424 , 12574501 , 12574524 , 12574556 , 12574599 , 12574619 , 12574731 , 12574749 , 12574853 , 12574858 , 12574860 , 12574865 , 12574961 , 12575626 , 12575698 , 12575886 , 12576125 , 12576443 , 12576485 , 12576518 , 12576593 , 12576609 , 12576710 , 12576722 , 12577271 , 12577685 , 12578856 , 12579183
- Product
- Plum A+ Infusion Pump; List Number: 11973; Product Usage: The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise…
- Recalling firm
- Hospira Inc.
- Quantity
- List No. 11973: 71 units
- Distribution
- Worldwide Distribution - USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, District of Columbia, Puerto Rico and U.S. Virgin Islands; and countries of: Croatia, Dominican Republic, Turks and Caicos.
- Recall date
- November 20, 2013
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)