FDAClass INovember 15, 2023

Evis Exera III Bronchovideoscope recall

EVIS EXERA III BRONCHOVIDEOSCOPE Model Numbers BF-1TH190, BF-H190, BF-Q190 and BRONCHOVIDEOSCOPE OLYMPUS BF-Q170 Model Number BF-Q170

Why was Evis Exera III Bronchovideoscope recalled?

There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model BF-1TH190 UDI-DI: 4953170335181; Model BF-H190 UDI-DI: 4953170335174; Model BF-Q170 UDI-DI: 4953170342912 Model BF-Q190 UDI-DI: 4953170335198

Product
EVIS EXERA III BRONCHOVIDEOSCOPE Model Numbers BF-1TH190, BF-H190, BF-Q190 and BRONCHOVIDEOSCOPE OLYMPUS BF-Q170 Model Number BF-Q170
Recalling firm
Olympus Corporation of the Americas
Quantity
32211 devices
Distribution
US Nationwide distribution
Recall date
November 15, 2023
Classification
Class I
Status
Ongoing
Agency
FDA (device)