Evis Exera II Bronchovideoscope Olympus BF Type recall
EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Numbers BF-1T180, BF-1TQ180, BF-P180
Why was Evis Exera II Bronchovideoscope Olympus BF Type recalled?
There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model BF-1T180 UDI-DI: N/A; Model BF-1TQ180 UDI-DI: N/A; Model BF-P180 UDI-DI: N/A;
- Product
- EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Numbers BF-1T180, BF-1TQ180, BF-P180
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 13950 devices
- Distribution
- US Nationwide distribution
- Recall date
- November 15, 2023
- Classification
- Class I
- Status
- Ongoing
- Agency
- FDA (device)