FDAClass IINovember 10, 2021

Arietta 850 Ultrasound- Software Version: V1.0.0 recall: Software Issue

Arietta 850 Ultrasound- Software Version: V1.0.0 through V4.1.3

Why was Arietta 850 Ultrasound- Software Version: V1.0.0 recalled?

Diagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display the measurement results correctly with two cardiology measurement functions: Mitral Regurgitation Flow (MR Flow) Measurement; Mitral Valve (MV) Measurement. In addition there is an error in the instruction manual.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Numbers: 205W3386 (01)04562122958974(11)171114(21)205W3386; 206F7900 (01)04562122958974(11)180926(21)206F7900; 206F7901 (01)04562122958974(11)180926(21)206F7901; 206F7936 (01)04562122958974(11)181113(21)206F7936; 206F7937 (01)04562122958974(11)181113(21)206F7937; 206F7940 (01)04562122958974(11)181113(21)206F7940; 206F7941 (01)04562122958974(11)181113(21)206F7941; 206F7942 (01)04562122958974(11)181113(21)206F7942; 206F7944 (01)04562122958974(11)181113(21)206F7944; 206F7945 (01)04562122958974(11)181113(21)206F7945; 206F7946 (01)04562122958974(11)181113(21)206F7946; 206F7947 (01)04562122958974(11)181114(21)206F7947; G3002552 (01)04562122958974(11)20190115(21)G3002552; G3024545 (01)04562122958974(11)190607(21)G3024545; G3037885 (01)04562122958974(11)20190719(21)G3037885; G3037887 (01)04562122958974(11)20190719(21)G3037887; G3042405 (01)04562122958974(11)190829(21)G3042405; G3042823 (01)04562122958974(11)20190807(21)G3042823; G3042828 (01)04562122958974(11)20190808(21)G3042828; G3045094 (01)04562122958974(11)20190822(21)G3045094; G3052105 (01)04562122958974(11)20191008(21)G3052105 G3052118 (01)04562122958974(11)20191025(21)G3052118 G3055549 (01)04562122958974(11)20191118(21)G3055549 G3055550 (01)04562122958974(11)20191118(21)G3055550 G3055552 (01)04562122958974(11)20191118(21)G3055552 G3055553 (01)04562122958974(11)20191118(21)G3055553 G3055554 (01)04562122958974(11)20191118(21)G3055554 G3055555 (01)04562122958974(11)20191118(21)G3055555 G3060234 (01)04562122958974(11)20191206(21)G3060234 G3060254 (01)04562122958974(11)20191206(21)G3060254 G3060275 (01)04562122958974(11)20191216(21)G3060275 G3060278 (01)04562122958974(11)20191216(21)G3060278 G3060690 (01)04562122958974(11)200127(21)G3060690 G3071074 (01)04562122958974(11)20200227(21)G3071074 G3080785 (01)04562122958974(11)201105(21)G3080785 G3090450 (01)04562122958974(11)201225(21)G3090450 G3110884 (01)04562122958974(11)210401(21)G3110884

Product
Arietta 850 Ultrasound- Software Version: V1.0.0 through V4.1.3
Recalling firm
FujiFilm Healthcare Americas Corporation
Quantity
38 units
Distribution
Nationwide
Recall date
November 10, 2021
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls