Flexor Check-Flo Introducer - Raabe Modification recall
Flexor Check-Flo Introducer - Raabe Modification, Introducer Set, RPN KCFW-5.0-38-90-RB-RAABE, GPN G13069 The introducer sheath incorporates a…
Why was Flexor Check-Flo Introducer - Raabe Modification recalled?
Affected lots may be manufactured incorrectly with the radiopaque marker band located just below the Check-Flo proximal fitting instead of the distal tip. The issue may not be identified until the device is placed in the patient and viewed under fluoroscopy, leading to increased procedural time.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
lot 13846597
- Product
- Flexor Check-Flo Introducer - Raabe Modification, Introducer Set, RPN KCFW-5.0-38-90-RB-RAABE, GPN G13069 The introducer sheath incorporates a…
- Recalling firm
- Cook Inc.
- Quantity
- 48 (US)
- Distribution
- Domestic distribution to AR AZ CT FL GA IL LA MA ME MN ND NJ NY OH PA TX WA. Foreign distribution to Germany, Finland, Italy, Netherlands, and Switzerland.
- Recall date
- November 10, 2021
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)