FDAClass IIOctober 30, 2019

Torrent Suite Dx Software Version recall: Software Issue

Torrent Suite Dx Software Version (IUO) 5.2 Catalog Number: A29225 OEM Customer Software: Sentosa SQ Suite Software v5.6.9

Why was Torrent Suite Dx Software Version recalled?

Torrent SuiteTM Dx analysis software, a component of the Ion PGM Dx Instrument System, incorrectly displays a positive result when a different analytical or de novo variant is detected at the same locus as a specified clinical variant

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

The field action is associated with the software available on the PGM Dx instrument, therefore no lot code is applicable for this part type

Product
Torrent Suite Dx Software Version (IUO) 5.2 Catalog Number: A29225 OEM Customer Software: Sentosa SQ Suite Software v5.6.9
Recalling firm
Life Technologies Holdings Pte Ltd
Quantity
total of Ion+vela +chassis =400 units
Distribution
states : AZ, CA, CO, CT, FL, GA, IN, MA, MD, NC, NJ, NM, NY, OR, TX, VA, WA OUS: South Korea, ITALY, China, Japan, Singapore, Germany, TAIWAN, Poland
Recall date
October 30, 2019
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls