FDAClass IIOctober 22, 2025

Medical convenience kits recall

Medical convenience kits Item Number/Description LKRA40G RADIOLOGY PACK

Why was Medical convenience kits recalled?

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

LKRA40G UDI-DI 191072189691 Lot 8025211 Bag serial number 68154347 * If bag serial number is listed as "N/A", then whole lot is affected

Product
Medical convenience kits Item Number/Description LKRA40G RADIOLOGY PACK
Recalling firm
American Contract Systems Inc.
Quantity
32,433 total
Distribution
US Nationwide distribution in the states of SD, IA, MN, WA, IL.
Recall date
October 22, 2025
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls