FDAClass IIDecember 6, 2017

Dimension Vista Protein 1 Control H recall

Dimension Vista Protein 1 Control H, PROT 1 CON H, Siemens Material Number (SMN) 10445921, Product Code KC717; IVD.

Why was Dimension Vista Protein 1 Control H recalled?

Complaint of high amount of ceruloplasmin patient results below the reference range (< 20 mg/dL [0.2 g/L]. PROT1 CAL and corresponding PROT1 CON L, M, and H are inaccurately assigned for ceruloplasmin (CER).

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot numbers: 6EQH01, 6EQH01A, 6EQH01B, 6EQH01C (exp. date 2017-09-17); 6LQH01A, 6LQH01B, 6LQH01C (exp. date 2018-04-04). --- Lot # (UDI): 6EQH01 (008427680063576EQH0120170917), 6EQH01A (008427680063576EQH01A2017091), 6EQH01B(008427680063576EQH01B20170917), 6EQH01C(008427680063576EQH01C20170917), 6LQH01A (008427680063576LQH01A20180404), 6LQH01B (008427680063576LQH01B20180404), 6LQH01C (008427680063576LQH01C20180404).

Product
Dimension Vista Protein 1 Control H, PROT 1 CON H, Siemens Material Number (SMN) 10445921, Product Code KC717; IVD.
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
N/A
Distribution
Worldwide Distribution - US (nationwide) Internationally to Austria, Australia, Belgium, Canada, France, Germany, Great Britain, Italy, Japan, Netherlands, Portugal, Slovakia, Spain, Switzerland and Republic of Korea South
Recall date
December 6, 2017
Classification
Class II
Status
Terminated
Agency
FDA (device)