FDAClass IINovember 3, 2021

Concerto User Interface Software provided with the recall

Concerto User Interface Software provided with the following systems: (1) Xstrahl 100 Electronic Brachytherapy - 100kV Superficial X-Ray Therapy…

Why was Concerto User Interface Software provided with the recalled?

If a saved treatment plan with 2 opposing beams is edited prior to approval, then Beam 2 is not updated with the changed parameters upon selecting save, resulting in error messages during the treatment and possible mis-treatment.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

V2.0, V2.1, and V2.2 when 2 opposing beam treatment plans are used.

Product
Concerto User Interface Software provided with the following systems: (1) Xstrahl 100 Electronic Brachytherapy - 100kV Superficial X-Ray Therapy…
Recalling firm
Xstrahl Limited
Quantity
25 systems in the U.S.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, FL, KY, NV, OR, PA, TN, and TX. The countries of Austria, Germany, and Switzerland.
Recall date
November 3, 2021
Classification
Class II
Status
Terminated
Agency
FDA (device)