FDAClass IINovember 9, 2022

BD Multitest 6-Color TBNK with optional BD Trucount recall

BD Multitest 6-Color TBNK with optional BD Trucount Tubes (Cat. No. 662995), for use with BD FACSLyric, BD FACSCanto II, and BD FACSCanto flow…

Why was BD Multitest 6-Color TBNK with optional BD Trucount recalled?

Due to partial label detachment on tubes. This can cause tubes to become stuck together in the pouch or cause automation errors when used with the automated loader for the wash assistant.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Corrected as of 11/04/2022: Catalog Number: 662995 UDI-DI Code: 00382906629951 Lot Numbers: 11568 45905 53637 88003 89111 31905

Product
BD Multitest 6-Color TBNK with optional BD Trucount Tubes (Cat. No. 662995), for use with BD FACSLyric, BD FACSCanto II, and BD FACSCanto flow…
Recalling firm
Becton, Dickinson and Company, BD Biosciences
Quantity
1,457 tubes (corrected on 11/04/2022)
Distribution
U.S.: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, VT, WA, and WI O.U.S.: Argentina, Australia, Belgium, Bolivia, Brazil, Canada, Chile, China, Columbia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Hong Kong, India, Japan, Korea, Mexico, Malaysia, New Zealand, Panama, Peru, Philippines, Paraguay, Singapore, Thailand, Taiwan, and Uruguay
Recall date
November 9, 2022
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls