FDAClass IIDecember 6, 2017

BD Phoenix PMIC-107 recall

BD Phoenix PMIC-107, Catalog Number 448417, UDI #s (01)30382904484171, (10)7031818(17)180228(30)0025, IVD. PMIC-107 is a Gram positive AST Only panel…

Why was BD Phoenix PMIC-107 recalled?

BD has recently confirmed quality control failures, related to a manufacturing issue, for the inducible macrolide-linosamide-streptogramin B (iMLSb) test contained on lot 7031818 of Phoenix panels. The iMLSb resistance D-test screen for Staphylococcus species is used to identify phenotypic drug resistance to clindamycin and erythromycin. As the product should not be used when quality control (QC) fails, it is unlikely that there would be any impact on patient results, however, if a customer does not perform QC testing and uses the panel, there is a potential to incorrectly treat a Staphylococcus infection with clindamycin based on the iMLSb test result. This issue can be detected 100% of the time when used per package insert instructions with iMLSb positive strain S. aureus BAA-977.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot 7031818, Exp. Date 2/28/2018

Product
BD Phoenix PMIC-107, Catalog Number 448417, UDI #s (01)30382904484171, (10)7031818(17)180228(30)0025, IVD. PMIC-107 is a Gram positive AST Only panel…
Recalling firm
Becton Dickinson & Co.
Quantity
US: 7750 units
Distribution
CT, VA, PA, TX, FL, NC, WI, NM
Recall date
December 6, 2017
Classification
Class II
Status
Terminated
Agency
FDA (device)