COBAS INTEGRA 800 Analyzer with software 9864.C2 Cobas. recall: Software Issue
COBAS INTEGRA 800 Analyzer with software 9864.C2 Cobas. COBAS Part Numbers 28122474001, 28122474692, 04559126001, 04589459970 corresponding to the…
Why was COBAS INTEGRA 800 Analyzer with software 9864.C2 Cobas. recalled?
Roche COBAS INTEGRA 800 system, Software Version 9864.C2 does not perform necessary ISE Service Actions. Whenever the ISE rack is removed, the ISE counters in the software are reset to zero and the corresponding Service Actions are not conducted automatically by the system. Additionally, certain maintenance activities that have to be performed manually by the operator are not requested on the sof
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
software version 9864.C2
- Product
- COBAS INTEGRA 800 Analyzer with software 9864.C2 Cobas. COBAS Part Numbers 28122474001, 28122474692, 04559126001, 04589459970 corresponding to the…
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Quantity
- 527 units
- Distribution
- Nationwide Distribution and Puerto Rico
- Recall date
- November 13, 2013
- Classification
- Class II
- Status
- Terminated