FDAClass IINovember 4, 2015

BG cartridges for IRMA TRUPOINT Blood Gas Analyzer recall

BG cartridges for IRMA TRUPOINT Blood Gas Analyzer, PN 048103. BG Cartridge are single use, disposable, in vitro diagnostics intended for point of…

Why was BG cartridges for IRMA TRUPOINT Blood Gas Analyzer recalled?

May report values outside the published performance specifications for pH. After a certain point in shelf life, the pH sensor may give results that are falsely elevated. The discrepancy is greater as shelf life progresses and can be seen on the aqueous controls before the blood.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

EOSCU EOTUZ EOVMW EOYUL EPANW EPCED EPDVM EPFOJ EPIXX EPKOJ EPNXA

Product
BG cartridges for IRMA TRUPOINT Blood Gas Analyzer, PN 048103. BG Cartridge are single use, disposable, in vitro diagnostics intended for point of…
Recalling firm
LifeHealth, LLC
Quantity
1735 boxes of 25. (43375 Cartridges)
Distribution
US: Nationwide (AL, AR, CA, CO, FL, IL, IN, IA, KS, KY, MD, MA,MI, MN, MS, MO, NE, NJ, NY,NC, OH, OK, PA, TN, TX, VA) OUS: ALBANIA, BELGIUM, BOLIVIA, CANADA, CHILE, CHINA, DENMARK. DOMINICAN REPUBLIC, ECUADOR, EGYPT, ENGLAND, ESTONIA, PHILIPPINES, FRANCE, GERMANY, GREECE, ICELAND, INDONESIA, IRELAND, ITALY, JAMAICA, JAPAN, MOROCCO, NETHERLANDS, NORWAY, PERU, PORTUGAL, SOUTH AFRICA, SOUTH KOREA, SPAIN, TAIWAN, TUNISIA, TURKEY, UNITED ARAB EMIRATES, VIETNAM.
Recall date
November 4, 2015
Classification
Class II
Status
Terminated
Agency
FDA (device)