FDAClass IIOctober 22, 2025

Airvo 2 recall

Airvo 2, REF: PT101US, and myAirvo 2, REF: PT100US, used with Disinfection Kit User Manual, REF: UI-185043723

Why was Airvo 2 recalled?

When unintentionally disconnected from power source, humidifier device (used to deliver high flow warmed and humidified respiratory gases) may sound audible alert for less than 120 seconds, which may delay user awareness to reconnect power, which could lead to oxygen desaturation, hypoxia. Disinfection kit user manual updated to include checking the audible alert prior to use on a new patient.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

REF/UDI-DI: PT101US/9420012422347; PT100US/9420012422248. Disinfection Kit User Manual all revisions before UI-185043723 rev P

Product
Airvo 2, REF: PT101US, and myAirvo 2, REF: PT100US, used with Disinfection Kit User Manual, REF: UI-185043723
Recalling firm
Fisher & Paykel Healthcare, Ltd.
Quantity
51,144
Distribution
US Nationwide distribution including in the states of TX, MS, LA, NC, TN, NY, FL, IL, UT, MT, OH, MI, NJ, AR, GA, WI, CA, IN, OK, CO, AL, SC, AZ, MD, WA, MO, PA, MA, NH, KY, KS, VA, CT, MN, ND, NM, NE, IA, WV, ID, NV, OR, HI, VT, RI, AL, SD, WY, PR, ME.
Recall date
October 22, 2025
Classification
Class II
Status
Ongoing
Agency
FDA (device)