OneTray Sealed Sterilization Containers Product Usage recall
OneTray Sealed Sterilization Containers Product Usage: A rigid sterilization container used to steam sterilize medical devices
Why was OneTray Sealed Sterilization Containers Product Usage recalled?
The wrong sterilization time was on the label. The label lists Steam Gravity 10 - 34 minutes at 270 degrees F instead of Steam Gravity 17 - 34 minutes at 270 degrees F.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
M2104 Lot # 044305, 044379, 044380 & 044418. M2408 Lot # 044296, 022304, 044381, 044420, 044421 & 044422.
- Product
- OneTray Sealed Sterilization Containers Product Usage: A rigid sterilization container used to steam sterilize medical devices
- Recalling firm
- Innovative Sterlization Technologies LLC
- Quantity
- M2104 - 25 units; M2408 - 84 units
- Distribution
- US Nationwide Distribution in the states of IL, IN, LA, OH, PA, TX & WV.
- Recall date
- October 28, 2015
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)