FDAClass IINovember 13, 2013

Adjustable Gastric Band recall

Adjustable Gastric Band, packaged alone in transparent PETG inner and outer blisters with Tyvek lids. Product Usage: Intended for use in surgical…

Why was Adjustable Gastric Band recalled?

During aging studies, the firm determined that the gastric belts were unable to pass the test requirements established for Force to Lock.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

The following product code and lot numbers were affected by this recall: Product Codes: 2200-X Lot Numbers: ZLGBC5, ZLNBCW, ZLNBJC, ZLPBB9, ZMBBBB, ZMCBBC, ZMGBCJ, ZMHBBT, ZMHBCV, ZMLBD0, ZMMBCD, ZMPBBC, ZNCBC6, ZNGBB3, and ZNGBBF.

Product
Adjustable Gastric Band, packaged alone in transparent PETG inner and outer blisters with Tyvek lids. Product Usage: Intended for use in surgical…
Recalling firm
Ethicon Endo-Surgery Inc
Quantity
Product Code: 2200-X - 2,475 units
Distribution
Worldwide Distribution: USA (nationwide) including the states of: Alabama, Connecticut, Indiana, Michigan, North Carolina, South Carolina, Alaska, Delaware, Kansas, Minnesota, Ohio, Tennessee, Arizona, Florida, Kentucky, Mississipp,i Oklahoma, Texas, Arkansas, Georgia, Louisiana, Nebraska, Oregon, Virginia, California, Idaho, Maryland, Nevada, Pennsylvania, Washington, Colorado, Illinois, Massachusetts, New Jersey, Rhode Island, and West Virginia; and the countries of: Argentina, France, Malaysia, Spain, Australia, Georgia, Martinique, Sweden, Belgium, Germany, Mexico, Switzerland, Brazil, Greece, Netherland, Taiwan, Chile, Hungary, Panama, Tunisia, China, India, Philippines, United Arab Emirates, Czech Republic, Israel, Poland, United Kingdom, Egypt, Italy, Portugal, Venezuela, Estonia, Kenya, Russia, Finland, Kuwait, and Saudi Arabia.
Recall date
November 13, 2013
Classification
Class II
Status
Terminated
Agency
FDA (device)