Freedom EVO-2 100 Base Unit recall
Freedom EVO-2 100 Base Unit, with Air LiHa Arms The Freedom EVO is an open automation platform product for general laboratory use. It is intended for…
Why was Freedom EVO-2 100 Base Unit recalled?
Distortion (noise interference) of the liquid level capacitance can result in air aspiration instead of liquid aspiration. This might cause a false signal (air detection) with possible dispense errors resulting in erroneous results.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Part Number: 10641100, SN: 1407006311.
- Product
- Freedom EVO-2 100 Base Unit, with Air LiHa Arms The Freedom EVO is an open automation platform product for general laboratory use. It is intended for…
- Recalling firm
- Tecan US, Inc.
- Quantity
- 2 units
- Distribution
- US Distribution to IL.
- Recall date
- November 5, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)