FDAClass IINovember 5, 2014

ACCU-CHEK Spirit Combo insulin infusion pump sold as recall

ACCU-CHEK Spirit Combo insulin infusion pump sold as part of Accu-Chek Combo System Product Usage: The ACCU-CHEK Combo System is meant only for…

Why was ACCU-CHEK Spirit Combo insulin infusion pump sold as recalled?

The insulin pump may lose time and date settings during a power interruption (such as a battery change) due to a faulty capacitor. If the capacitor fails, the time and date will return to default setting. Though the pump prompts customers to confirm the time and date, if the user does not recognize the time and date have returned to default, a shift of their basal rate time block could occur.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model/Catalog/Part Number: Accu-Chek Combo Kit mg US/English version, 05458544001 Accu-Chek Combo Kit mg US/Spanish version, 055004686001 Accu-Chek Spirit Combo Replacement Pump US/EN, 05391512001 Accu-Chek Spirit Combo Demo Pump- US, 0539175000

Product
ACCU-CHEK Spirit Combo insulin infusion pump sold as part of Accu-Chek Combo System Product Usage: The ACCU-CHEK Combo System is meant only for…
Recalling firm
Roche Diagnostics Operations, Inc.
Quantity
8116
Distribution
US Nationwide Distribution in the states of CA, CT, IA, NV, OH, IN, WI, GA, MO, NY, VA, LA, ME, TX, NC, ND, SD, NJ, CO, AL, SC, IL, KY, FL, OR, MN, MD, MS, AR, WV, WA, TN, CT, ID, KS, MI, PA, UT, MA, RI, NM, OK, DE, AZ, NE, MT, NV, WY, AK, and NH.
Recall date
November 5, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls