Bayer Brand A1cnow+ Multi-test A1c System Test Kits recall
Bayer brand A1CNow+ MULTI-TEST A1C SYSTEM TEST KITS (PART NUMBER 3021/08842610); Hemoglobin A1c Monitor; Glycosylated Hemoglobin Assay; Product is manufactured and distributed by Bayer HealthCare LLC, Sunnyvale, CA The A1CNow+ test provides quantitative measurement of the percent of glycated hemoglobin (%HbA1c or %A1C) levels in capillary (fingerstick) or venous whole blood samples. The test is for professional use to monitor glycemic control in people with diabetes.
Why was Bayer Brand A1cnow+ Multi-test A1c System Test Kits recalled?
Test results from these lots may produce a high-bias reading (by up to 20%) that is outside of limits defined for stability and product accuracy when product is stored at room temperature four (4) to five (5) months after release.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot: 1300733 (US, China); Lot: 1300712 (Argentina, Uruguay); Lots: 1300711, 130012 (Japan)
- Product
- Bayer brand A1CNow+ MULTI-TEST A1C SYSTEM TEST KITS (PART NUMBER 3021/08842610); Hemoglobin A1c Monitor; Glycosylated Hemoglobin Assay; Product is manufactured and distributed by Bayer HealthCare LLC, Sunnyvale, CA The A1CNow+ test provides quantitative measurement of the percent of glycated hemoglobin (%HbA1c or %A1C) levels in capillary (fingerstick) or venous whole blood samples. The test is for professional use to monitor glycemic control in people with diabetes.
- Recalling firm
- Bayer Healthcare Llc
- Quantity
- 138 Kits in US (Total 2452)
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to China, Argentina, Uruguay, Japan.
- Recall date
- November 13, 2013
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)