FDAClass IIOctober 22, 2025

Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation recall

Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0740-140-2…

Why was Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation recalled?

Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter that may not open or may not fully open during sheath retraction and operation of the filter knob. The stent is used with a primary distal embolic protection device.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

REF/UDI-DI/Lot(Expiration): NG-0740-140-2/00851616007613/Z2460707D(Z2460707D), Z2460733A(03/05/2027), Z2460766A(04/07/2027), Z2460774B(04/17/2027), Z2460780A(04/18/2027); NG-0640-140-2/00851616007590/Z2460746C(03/20/2027), Z2460774A(04/18/2027); NG-NV-7-40/00851616007620/Z2460776D(04/14/2027); NG-NV-6-40/00851616007651/Z2460776B(04/16/2027)

Product
Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0740-140-2…
Recalling firm
Contego Medical LLC
Quantity
568
Distribution
US Nationwide distribution in the states of FL, CO, CA, IL, NC, KY, TX, PA, DE, NJ, IN, NY, OH, LA, OK, SC, CT, WI, MA, KS, UT, MD, AL.
Recall date
October 22, 2025
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls