Brand Name: NOVA ORTHO-MED INC. Product Name: MONARCH recall
Brand Name: NOVA ORTHO-MED INC. Product Name: MONARCH ROLLATOR RED Model/Catalog Number: 4329RD Product Description: ROLLING WALKER MOBILITY AID
Why was Brand Name: NOVA ORTHO-MED INC. Product Name: MONARCH recalled?
Firm identified a potential risk for a fall hazard. When the end user is readjusting from the slouched seated position to an upright seated position, there is the potential for the movement to cause the backrest to lift out of the backrest holder and disengage from the unit.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot Code: ITEM CODE: 4329RD SN RANGE: HT4329RDME0001 - HT4329RDMM0420 UPC: 652308012263
- Product
- Brand Name: NOVA ORTHO-MED INC. Product Name: MONARCH ROLLATOR RED Model/Catalog Number: 4329RD Product Description: ROLLING WALKER MOBILITY AID
- Recalling firm
- Nova Ortho-Med Inc
- Quantity
- 1585
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AK, AL AZ, CA, CO, CT, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, WA, WI and WY. The country of Canada.
- Recall date
- October 23, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)