Carestream DRX-Revolution Mobile X-ray System. Produces recall
Carestream DRX-Revolution Mobile X-ray System. Produces radiographic images.
Why was Carestream DRX-Revolution Mobile X-ray System. Produces recalled?
Carestream Health Inc. is recalling the power plug on their DRX-Revolution Mobile X-Ray System due to the item generating sparks and becoming warm to touch.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Serial Numbers: 111, 114, 119-123, 125-126, 134, 142-143, 151-155, 161-169, 173-178,180-183, 188-192, 196-209, 211, 221-232, 235-249, 254, 261-270, 275-276, 279-286, 305, 312, 316-326, 330-337, 339-349, 353, 355, 357-365, 372-373, 375-376, 379-385, 387-388, 394-401, 403-405, 409-410, 412-415, 417, 420-421, 423, 431-435, 441-446, 451-461, 464-470, and 482-490
- Product
- Carestream DRX-Revolution Mobile X-ray System. Produces radiographic images.
- Recalling firm
- Carestream Health, Inc.
- Quantity
- 180 units
- Distribution
- Nationwide Distribution including the states of NY, TX, PA, OH, WI, CA, IL, CO, FL, IN, OK, NJ, GA, LA, AL, KS, AZ, MA, SC, MI, NC, IA, UT, VA, TN, and MN.
- Recall date
- November 6, 2013
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)