Elekta Medical Linear Accelerator recall
Elekta Medical Linear Accelerator (EMLA): Elekta Harmony Pro, REF 1555143; Elekta Infinity, REF 1552091, and Elekta Versa HD, REF 1552093.
Why was Elekta Medical Linear Accelerator recalled?
There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Elekta Harmony Pro, Serial #441009, UDI (01) 05060191071604; Elekta Infinity, Serial #156645, UDI (01) 05060191071543; and Elekta Versa HD, Serial #156646, #156477, and #156648, UDI (01) 05060191071574
- Product
- Elekta Medical Linear Accelerator (EMLA): Elekta Harmony Pro, REF 1555143; Elekta Infinity, REF 1552091, and Elekta Versa HD, REF 1552093.
- Recalling firm
- ELEKTA SOLUTIONS AB
- Quantity
- 2 in the U.S.
- Distribution
- Distribution was to ND and NC. There was no military/government distribution.
- Recall date
- October 20, 2021
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)