Astral 100 and Astral 150 ventilators recall
Astral 100 and Astral 150 ventilators
Why was Astral 100 and Astral 150 ventilators recalled?
If ventilator is on internal battery, not intended to serve as a primary power source, low/critically low battery alarms will sound, but a fault leads to sudden power loss. If power fails, then ventilation stops and a Total Power Failure (TPF) alarm should sound, but it's powered by a supercapacitor, which degrades over time, which may cause TPF alarm to sound for less than 2 minutes or not at all
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All devices manufactured between 2013 and 2019. Device label lists the serial number, which includes the year of manufacture. For example, if the serial number is 22151234567, the 3rd and 4th digit are 15, meaning the device was manufactured in 2015. Astral 100, UDI: 00619498003259 Astral 150, UDI: 00619498003266
- Product
- Astral 100 and Astral 150 ventilators
- Recalling firm
- ResMed Ltd.
- Quantity
- 55,279 (US 16,634; OUS 38,645)
- Distribution
- Worldwide - US Nationwide distribution including in the states of CA, NJ, VA, IN, MI, WV, NY, WI, ND, IL, NC, FL, CO, TX, AL, LA, MD, OR, AZ, MN, WA, PA, SC, MA, ID, GA, SD, TN, NH, OH, MS, CT, IA, KY, UT, NE, MO, NV, AR, WY, RI, KS, VT, ME, OK, PR, MO, DC and the countries of Taiwan, Israel, Spain, Slovenia, Belgium, Italy, Australia, Guatemala, Argentina, Uruguay, Chile, France, Martinique, Germany, French Polynesia, Korea, Egypt, Colombia, United Kingdom, Jersey, Ecuador, Reunion, Greece, Brazil, Cyprus, Saudi Arabia, Czech Republic, Thailand, New Caledonia, Mauritius, Canada, Vietnam, Indonesia, Pakistan, Japan, Guadeloupe, South Africa, Iran, Hungary, Tunisia, Nepal, Iceland, New Zealand, Austria, Portugal, Malaysia, Norway, Netherlands, Sweden, Lebanon, Finland, Singapore, United Arab Emirates, Peru, Oman, Philippines, Bahrain, French Guiana, Mexico, Sri Lanka, India, Kuwait, Bangladesh, Kenya, Myanmar, Turkey, Bulgaria, Croatia, Hong Kong, Saint Martin, Qatar, Jordan, Romania, Maldives, Denmark, Poland.
- Recall date
- November 1, 2023
- Classification
- Class I
- Status
- Ongoing
- Agency
- FDA (device)