YUKON Polyaxial Screws recall
YUKON Polyaxial Screws (3.5 X 12 mm, Owning Site: Leesburg USA 3.5 X 14 mm, 3.5 X 32 mm, 3.5 X 34 mm, 3.5 X 38 mm, 3.5 X 40 mm, 4.0 X 16 mm, 4.0 X 18…
Why was YUKON Polyaxial Screws recalled?
Screws had a manufacturing error that could prevent the screw from mating with the screw inserter or set screw as intended. This event was reported to FDA as part of Stryker's retrospective review of K2M product field actions; the action was completed in 2018.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lots FUVD, FUVE, FUVG, FUWF, FUWK, FUWL, FUWU, FUWV, FUWW, FUXK, FUXL,FUXM, FUXN, FUXP, FUXR, FUXT, FUXU, FUXV
- Product
- YUKON Polyaxial Screws (3.5 X 12 mm, Owning Site: Leesburg USA 3.5 X 14 mm, 3.5 X 32 mm, 3.5 X 34 mm, 3.5 X 38 mm, 3.5 X 40 mm, 4.0 X 16 mm, 4.0 X 18…
- Recalling firm
- K2M, Inc
- Quantity
- 364
- Distribution
- Distributed to accounts in AL, CA, CO, FL, IL, KY, LA, MD, MI, MN, MS, NC, NJ, NY, OH, OK, PA, TN, TX, and WA.. Foreign distribution to Australia.
- Recall date
- October 16, 2019
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)