Multichem P recall
Multichem P, Part Code 08P90-10
Why was Multichem P recalled?
Technopath Manufacturing Ltd. has identified an issue with the glass vial commodity used to fill the Multichem S Plus Level 1, Multichem S Plus Level 2, Multichem S Plus Level 3, Multichem IA Plus and Multichem P Controls. Some customers have experienced broken vials when receiving or thawing Technopath Multichem IA Plus and Multichem S Plus Controls. Breakages have been observed upon delivery, while thawing or while opening the vial. The breakages can result in the bottom of the vial breaking off, breakage in the neck thread area or cracking on the sides of the vial. This may expose the user to biohazardous material if cut and may cause a delay in patient results.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
UDI/DI 05391523440769, Lot Numbers: 04530621P, 04561021P, 04610523P
- Product
- Multichem P, Part Code 08P90-10
- Recalling firm
- TECHNO-PATH MANUFACTURING LTD.
- Quantity
- 2964 units
- Distribution
- US Nationwide distribution in the state of Illinois.
- Recall date
- October 18, 2023
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)