Ultra Battery 1 recall
Ultra Battery 1 Model: TA-UB1 NSN: 6515-01-592-5803 The Ultra Battery 1 (TA-UB1) is a rechargeable power source for the Thermal Angel Blood and IV…
Why was Ultra Battery 1 recalled?
Estill has discovered that a fully discharged Ultra Battery 1 (TA-UB1) will reach 96.9% of its capacity during a 60 minute charging period. If you have a fully discharged Ultra Battery 1 (TA-UB1), in order to achieve a 100% of charge capacity you should charge it for a period of 90 minutes. In addition, the charging voltage range of 10-36V specified on the Ultra Battery 1 (TA-UB1) label and in t
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
06-2011; 07-2011; 08-2011; 09-2011; 03-2012; 04-2012; 05-2012; 06-2012; 07-2012; 08-2012; 01-2013; 03-2013; 08-2013; 09-2013; 10-2013; 11-2013; 12-2013; 01-2014; 02-2014; 04-2014; 07-2012; 02-2014
- Product
- Ultra Battery 1 Model: TA-UB1 NSN: 6515-01-592-5803 The Ultra Battery 1 (TA-UB1) is a rechargeable power source for the Thermal Angel Blood and IV…
- Recalling firm
- Estill Medical Technologies, Inc
- Quantity
- 6,886
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Serbia.
- Recall date
- October 29, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)