FDAClass IIIOctober 23, 2024

Beckman Coulter IRISPEC CA/CB/CC Urine Chemistry recall

Beckman Coulter IRISPEC CA/CB/CC Urine Chemistry Controls, Catalog Number REF 800-7211 and 800-7702, in vitro diagnostic device.

Why was Beckman Coulter IRISPEC CA/CB/CC Urine Chemistry recalled?

Beckman Coulter has become aware of an increase in customer complaints for intermittent IRISpec CB negative glucose control failure, resulting in false positive control results when used in conjunction with iChemVELOCITY Urine Chemistry Strips (PNs 800-7204 and 800-7212) on the iChemVELOCITY Analyzer. This issue may delay the reporting of patient results, with a worst-case scenario of a delay in diagnosis or treatment of metabolic disorders, kidney function abnormalities, urinary tract infections, and liver function.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI/DI 10837461002611, Lot Numbers: 432-24 and higher

Product
Beckman Coulter IRISPEC CA/CB/CC Urine Chemistry Controls, Catalog Number REF 800-7211 and 800-7702, in vitro diagnostic device.
Recalling firm
Beckman Coulter, Inc.
Quantity
5589 units
Distribution
worldwide
Recall date
October 23, 2024
Classification
Class III
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls