FDAClass IINovember 6, 2013

DRX Revolution Mobile X-Ray System recall: Software Issue

DRX Revolution Mobile X-Ray System, Model DRXR-1, Verona Street Rochester, NY 14608 Catalog Numbers 1019397, 1023415, 1023423 --- Made in USA by…

Why was DRX Revolution Mobile X-Ray System recalled?

During an audit, Carestream discovered a DRX-1 System that captured an image which displayed an artifact. This issue was discovered in manufacturing. It occurs when the DRX product console software is updated to v5.6b software and the detector is not calibrated before use.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial numbers: U.S. - 134,152,173,174,200,205,222,312,349,353,412, Worldwide Serial numbers: 469, 445, 306, 410, 453, 204, 483, 300, 171, 218

Product
DRX Revolution Mobile X-Ray System, Model DRXR-1, Verona Street Rochester, NY 14608 Catalog Numbers 1019397, 1023415, 1023423 --- Made in USA by…
Recalling firm
Carestream Health, Inc.
Quantity
23 units installed total (11 domestically & 12 internationally)
Distribution
Worldwide Distribution - US (nationwide) including the states of New York, Massachusetts, Pennsylvania, Illinois, Texas, California and Louisiana., and the countries of Australia, Canada, Finland, Italy, Switzerland and United Kingdom.
Recall date
November 6, 2013
Classification
Class II
Status
Terminated
Agency
FDA (device)