Uniflex Humeral Nail recall
Uniflex Humeral Nail: 467265 G-PIN 2.8BL 3.2PL 60CM 2PK Pin, Fixation, Smooth A 467265-01 Ball Tip Guide Pin is used to guide reamers down the humeral…
Why was Uniflex Humeral Nail recalled?
The G-PIN 2.8BL 3.2PL 60 CM 2PK is intended to contain one (1) 467265-01 Ball Tip Guide Pin and one (1) 467255-02 Plain Tip Guide Pin. The G-PIN 2.8BL 3.2PL 60 CM 2PK may have been packaged incorrectly either with two (2) 467265-01 Ball Tip Guide Pins and zero (0) 467255-02 Plain Tip Guide Pins or with zero (0) 467265-01 Ball Tip Guide Pins and two (2) 467255-02 Plain Tip Guide Pins. If the G-PIN 2.8BL 3.2PL 60 CM 2PK contains the incorrect combination of Plain Tip Guide Pins and Ball Tip Guide Pins, the surgeon would either have to implant the nail without the Plain Tip Guide Pin as a guide, ream with a Plain Tip Guide Pin, or choose to not ream the humeral canal. The risk associated with not having the correct combination can include a delay in surgery of less than 30 minutes, the reamer tip remaining in the distal humeral canal, or an iatrogenic fracture.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Item Number 467265 Lot Number 662420
- Product
- Uniflex Humeral Nail: 467265 G-PIN 2.8BL 3.2PL 60CM 2PK Pin, Fixation, Smooth A 467265-01 Ball Tip Guide Pin is used to guide reamers down the humeral…
- Recalling firm
- Zimmer Biomet, Inc.
- Quantity
- 22 units
- Distribution
- US Distribution to states of: MD,TX, FL, GA, VA and PA.
- Recall date
- October 26, 2016
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)