Varian Head Frame recall
Varian Head Frame - Model number HHF -ALL Head Frame Posts (PN1008016 ) and Post Kits (PN1008016). The Varian Head Frame System is for use with a…
Why was Varian Head Frame recalled?
Varian Medical Systems has received a report that a user was able to easily rotate the head frame posts when attached to the mounting cam on the metal head ring. There was no report of serious injury due to this issue
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All Head Frame Posts and Post kits
- Product
- Varian Head Frame - Model number HHF -ALL Head Frame Posts (PN1008016 ) and Post Kits (PN1008016). The Varian Head Frame System is for use with a…
- Recalling firm
- Varian Medical Systems, Inc.
- Quantity
- 75 devices
- Distribution
- US Distribution and Internationally to 1 site in Canada, India and Morocco
- Recall date
- October 19, 2016
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)