"***Electric Systems Foot Control****Rx Only***" recall
"***Electric Systems Foot Control****Rx Only***" Product Usage: Cutting and shaping bone including bones of the spine and cranium.
Why was "***Electric Systems Foot Control****Rx Only***" recalled?
The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling the foot control/foot pedal due to the Foot Control Cable lacking the shield connection from the cable to the connector body.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Serial Numbers: E31305311301, E31305311302, E31305311303, E31305311304, E31305311401, E31305311402, E31305311403, E31305311404, E31305311405, E32305311201, E32305311202, E33305311501, E33305311502, E33305311504, and E33305311505
- Product
- "***Electric Systems Foot Control****Rx Only***" Product Usage: Cutting and shaping bone including bones of the spine and cranium.
- Recalling firm
- The Anspach Effort, Inc.
- Quantity
- 16 units
- Distribution
- Worldwide Distribution - US (nationwide) and the countries of: Italy, Japan, Switzerland, Turkey and Taiwan.
- Recall date
- October 24, 2012
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)