FDAClass IIOctober 24, 2012

Terumo Backplate Replacement CXFVR 1500 Holder The recall

Terumo Backplate Replacement CXFVR 1500 Holder The Flexible venous Reservoir is mounted on the Backplate/Holder and secures the flexible reservoir by…

Why was Terumo Backplate Replacement CXFVR 1500 Holder The recalled?

Backplate reservoir holders were manufactured with an incorrect configuration of connector supports. The devices were manufactured and configured with two 3/8 inch and one ¿ inch connector supports instead of one 3/8 inch and two ¿ inch connector supports (See location of supports as # 7 and # 8 on backplate attachment). The backplates were manufactured as described in manufacturing instructions a

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

No lot/serial number; Catalog number: PN 816280

Product
Terumo Backplate Replacement CXFVR 1500 Holder The Flexible venous Reservoir is mounted on the Backplate/Holder and secures the flexible reservoir by…
Recalling firm
Terumo Cardiovascular Systems Corporation
Quantity
33 units
Distribution
Worldwide distribution: USA (nationwide) including states of: CA, MI, NC, NJ, PA, and WI;and country of: EUROPE.
Recall date
October 24, 2012
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls