FDAClass IINovember 15, 2017

Delta Cup Beater  Positioner - Aligner recall

Delta Cup Beater  Positioner - Aligner, model # 9057.20.555 Used during implantation of the Delta Acetabular System

Why was Delta Cup Beater  Positioner - Aligner recalled?

An internal analysis on complaints received indicated that the welding process may have been performed sub optimally in the 12 lot # involved , leading to reduced mechanical strength of the device and possible breakage during impaction when the instrument is subjected to repeated multi-axial stresses.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot # 15AG0HJ, 15AG0HQ, 15AG0UY, 15AG00P, 15AG040, 15AG08Y, 14AG243, 15AG00Q, 15AG0UZ, 15AG0V0, 15AG0HN, 16AG02J

Product
Delta Cup Beater  Positioner - Aligner, model # 9057.20.555 Used during implantation of the Delta Acetabular System
Recalling firm
Limacorporate S.p.A
Quantity
20 units in (US)
Distribution
Worldwide Distribution - US Distribution and to the countries of : Australia, Austria, Belgium, Brazil, Czech Republic, Germany, Italy, New Zealand, Poland, Russia, Slovakia, Slovenia, Spain, South Africa, Switzerland, Turkey, UK and Germany
Recall date
November 15, 2017
Classification
Class II