FDAClass IIOctober 14, 2020

DJO EMPOWR Tibial Knee Punch Handle recall

DJO EMPOWR Tibial Knee Punch Handle Part No. 801-05-040 and 801-05-018 contained in Instrument Tray for EMPOWR Porous Knee - Product Usage: The Tibial…

Why was DJO EMPOWR Tibial Knee Punch Handle recalled?

There is a potential for the retaining pin of the tibial punch handle to become loose and fall out while impacting. If undetected, there is a potential of leaving debris in the joint space.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model/Catalog no. 801-05-040; :Lot/serial No.s 281905L08, 281905L15, 283808L01, 305197L02, 318871L10 Model/Catalog no. 801-05-018, Lot/serial No.s: 215488L08, 236760L01, 228231L01, 241946L02, 259710L01, 314969L01

Product
DJO EMPOWR Tibial Knee Punch Handle Part No. 801-05-040 and 801-05-018 contained in Instrument Tray for EMPOWR Porous Knee - Product Usage: The Tibial…
Recalling firm
Encore Medical, LP
Quantity
190 units
Distribution
U.S. Nationwide distribution including in the states of VA, TN, MN, TX, IL, IN, KY, OH, SC, RI, MD, PA, NY, CA, KS, OK, AR, LA, FL, AL, GA, AZ, NV, WA, MS, ME, NV and Puerto Rico.
Recall date
October 14, 2020
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls