KINOVA Jaco assistive robot recall
KINOVA Jaco assistive robot. Model Numbers: PJ 0000 0001, PJ 0000 0012, PJ 0090 0001, PJ 0090 0006, KR MJ2 0001. Jaco replaces the function of the…
Why was KINOVA Jaco assistive robot recalled?
Potential installation issue that could result in a fire hazard in the event of an interaction between a damaged Jaco arm and a wheelchair exhibiting electrical leakage.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model Numbers: PJ 0000 0001, PJ 0000 0012, PJ 0090 0001, PJ 0090 0006, KR MJ2 0001. All lot numbers.
- Product
- KINOVA Jaco assistive robot. Model Numbers: PJ 0000 0001, PJ 0000 0012, PJ 0090 0001, PJ 0090 0006, KR MJ2 0001. Jaco replaces the function of the…
- Recalling firm
- Kinova Inc
- Quantity
- 949 units
- Distribution
- Worldwide - US Nationwide distribution in the states of California, Missouri, North Carolina, Ohio and the countries of Australia, Canada, China, Czechia, France, Germany, Netherlands, Norway, Portugal, Switzerland, Great Britain, Northern Ireland.
- Recall date
- October 23, 2024
- Classification
- Class I
- Status
- Ongoing
- Agency
- FDA (device)