FDAClass IINovember 1, 2017

Power Processor 1K Stockyard. The Power Processor recall: Software Issue

Power Processor 1K Stockyard. The Power Processor performs all pre-analytical sample tube preparation.

Why was Power Processor 1K Stockyard. The Power Processor recalled?

Beckman Coulter has identified that due to a PLC software nonconformity the 1K Stockyard can initiate a retrieval of a sample tube during the rack loading process, which should not occur. This issue and associated complaint were discovered and filed internally.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Software version PLC V26R10

Product
Power Processor 1K Stockyard. The Power Processor performs all pre-analytical sample tube preparation.
Recalling firm
Beckman Coulter Inc.
Quantity
27 units
Distribution
China France Italy Kuwait Spain Turkey United States
Recall date
November 1, 2017
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls