FDAClass IIOctober 15, 2025

FDR Visionary Suite recall

FDR Visionary Suite, CH-200 Model/Version Numbers: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33

Why was FDR Visionary Suite recalled?

It was found that the shaft for mounting the X-ray tube unit on the support may break. In addition, it was found that the protective parts might not work effectively only when the shaft was broken while being operated toward the front while lifting.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model/Version Number: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33; UDI-DI: (1) 04540217052226, (2) 04540217057436, (3) 04540217057450; Serial Numbers: (1) MP95A8F5A001, MP95A9F6A001, MP95A9F6B001, MP95A9F74001, MP95A9F74002, MP95A9F76001, MP95A9F78001, (2) MP95AA783001, (3) MP95AA983001, MP95AA983002, MP95AA983003, MP95AA985001, MP95AA985003, MP95AA985002, MP95AA988001, MP95AA988002;

Product
FDR Visionary Suite, CH-200 Model/Version Numbers: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33
Recalling firm
FUJIFILM Healthcare Americas Corporation
Quantity
16 units
Distribution
Domestic: AK, CA, FL, IL, MA, NH, OH, WI.
Recall date
October 15, 2025
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls