FDAClass IIOctober 30, 2013

STERRAD 100NX Cassettes The STERRAD Sterilization recall

STERRAD 100NX Cassettes The STERRAD Sterilization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive…

Why was STERRAD 100NX Cassettes The STERRAD Sterilization recalled?

Advanced Sterilization Products (ASP) is recalling the three lots (13F049, 13F050, 13F051) STERRAD 100NX Cassettes because ASP has determined that the leak indicator on the cassettes is effective for only 12 months of the cassette's labeled 15-month shelf-life.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot# 13F050, 13F051, 13F049

Product
STERRAD 100NX Cassettes The STERRAD Sterilization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive…
Recalling firm
Advanced Sterilization Products
Quantity
2,995 units
Distribution
Nationwide Distribution
Recall date
October 30, 2013
Classification
Class II
Status
Terminated
Agency
FDA (device)