FDAClass IIOctober 13, 2021

Access hsTnI High Sensitivity Troponin I recall: Contamination

Access hsTnI High Sensitivity Troponin I, REF B52699 Access Immunoassay System REAGENT PACK

Why was Access hsTnI High Sensitivity Troponin I recalled?

Through customer feedback, testing a sample with cardiac troponin I concentration that are >270,000 pg/mL may cause carryover into the hsTnl reagent pack. The carryover is caused by sample-to-pack carryover predominantly into the particle well of the reagent pack by way of the probe. The particle well then becomes contaminated by the high sample from the probe. The extent of the carryover is dependent on sample concentration and frequency of high concentration samples.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Catalog Number:B52699 Lot Number / UDI Code: 921491 / (01)15099590693183(17)201130(11)191201(10)921491; 921638 / (01)15099590693183(17)201231(11)200101(10)921638; 921631 / (01)15099590693183(17)210131(11)200201(10)921631; 921750 / (01)15099590693183(17)210131(11)200201(10)921750; 921805 / (01)15099590693183(17)210228(11)200229(10)921805; 921831 / (01)15099590693183(17)210331(11)200331(10)921831; 921943 / (01)15099590693183(17)210430(11)200430(10)921943; 922000 / (01)15099590693183(17)210531(11)200531(10)922000; 922039 / (01)15099590693183(17)210630(11)200630(10)922039; 922040 / (01)15099590693183(17)210731(11)200731(10)922040; 922062 / (01)15099590693183(17)210731(11)200731(10)922062; 922145 / (01)15099590693183(17)210831(11)200831(10)922145; 922146 / (01)15099590693183(17)210930(11)200930(10)922146; 922338 / (01)15099590693183(17)210930(11)200930(10)922338; 922440 / (01)15099590693183(17)211031(11)201031(10)922440; 922545 / (01)15099590693183(17)211031(11)201031(10)922545; 922547 / (01)15099590693183(17)211130(11)201130(10)922547; 922713 / (01)15099590693183(17)211231(11)201231(10)922713; 922714 / (01)15099590693183(17)211231(11)201231(10)922714; 922722 / (01)15099590693183(17)220131(11)210131(10)922722; 922897 / (01)15099590693183(17)220131(11)210131(10)922897; 124235 / (01)15099590693183(17)220228(11)210228(10)124235; 124339 / (01)15099590693183(17)220228(11)210228(10)124339; 922898 / (01)15099590693183(17)220228(11)210228(10)922898; 124340 / (01)15099590693183(17)220331(11)210331(10)124340; 124348 / (01)15099590693183(17)220430(11)210430(10)124348; 124353 / (01)15099590693183(17)220430(11)210430(10)124353; 124688 / (01)15099590693183(17)220430(11)210430(10)124688; 124354 / (01)15099590693183(17)220531(11)210531(10)124354; 124355 / (01)15099590693183(17)220630(11)210630(10)124355; 124356 / (01)15099590693183(17)220630(11)210630(10)124356; 124357 / (01)15099590693183(17)220630(11)210630(10)124357;

Product
Access hsTnI High Sensitivity Troponin I, REF B52699 Access Immunoassay System REAGENT PACK
Recalling firm
Beckman Coulter Inc.
Quantity
293,640 kits
Distribution
U.S.: AL, AR, AZ, CA, CO, CT, DE, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, ME, MD, MI, MN, MS, MO, NC, ND, NE, NV, NJ, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, VA, WA, WI, WV, and WY O.U.S.: Algeria, Andorra, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, China, Croatia, Cura¿ao, Czech Republic, Denmark, Egypt, Estonia, Ethiopia, Finland, France, Georgia, Germany, Ghana, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Malaysia, Mayotte, Mexico, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Niger, Nigeria, Norway, Oman, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Swaziland, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Vietnam, Yemen, and Zimbabwe
Recall date
October 13, 2021
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls