FDAClass IIOctober 13, 2021

TECNIS Eyhance IOLs are permanent intraocular implants recall: Endotoxin

TECNIS Eyhance IOLs are permanent intraocular implants intended for implantation in the capsular bag. The TECNIS Simplicity Delivery System is…

Why was TECNIS Eyhance IOLs are permanent intraocular implants recalled?

Nonconforming product was distributed in error. Product was nonconforming due endotoxin levels higher than both the Firm's internal endotoxin specification limit and the FDA recommended endotoxin specification limit. Product was released in error based on the results of additional testing performed during the investigation into release testing nonconformance.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model: DIU225U050 UDI Code: (01)05050474745831(17)240414(21)5248142104 Production Order Number: 900001985625 Serial Number: 5248142104

Product
TECNIS Eyhance IOLs are permanent intraocular implants intended for implantation in the capsular bag. The TECNIS Simplicity Delivery System is…
Recalling firm
Johnson & Johnson Surgical Vision Inc
Quantity
1 lens
Distribution
U.S. Nationwide distribution in the state of IL.
Recall date
October 13, 2021
Classification
Class II
Status
Terminated
Agency
FDA (device)