FDAClass IIOctober 16, 2024

BD Synapsys Informatics Solution with software recall: Software Issue

BD Synapsys Informatics Solution with software versions: 4.20, 4.30, 5.10, 5.20 and 5.30- IVD standalone laboratory software solution providing data…

Why was BD Synapsys Informatics Solution with software recalled?

Synapsys software versions 4.20, 4.30, 5.10, 5.20, or 5.30 and utilizing the in-line breakpoint feature, an error is occurring where BD Synapsys will incorrectly display breakpoints for organisms that do not have defined breakpoints in CLSI or EUCAST. This occurs when the identified organism is included in the same interpretive group as a species for which there is a specific breakpoint. May result in either a false susceptible or false resistant result. A false susceptible result can cause inadequate or incomplete antibiotic treatment due to the use of an ineffective antibiotic, and a false resistant result may limit access to a preferred antibiotic regimen with possible elongation of therapy

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI: 00382904441500 Software versions: 4.20, 4.30, 5.10, 5.20 and 5.30

Product
BD Synapsys Informatics Solution with software versions: 4.20, 4.30, 5.10, 5.20 and 5.30- IVD standalone laboratory software solution providing data…
Recalling firm
Becton Dickinson & Co.
Quantity
162 units
Distribution
Nationwide Foreign: Australia, New Zealand, Singapore, Hong Kong, Taiwan, Japan, Austria, Belgium, Denmark, France, Germany, Italy, Luxembourg, Netherlands, Norway, Spain, Sweden, Switzerland, United Kingdom, Russia, Saudi Arabia, South Africa, Canada, Mexico, China
Recall date
October 16, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls