Ortho OS Solutions UltOS 3.5mm locking screw 12mm to recall
Ortho OS Solutions UltOS 3.5mm locking screw 12mm to 50mm Ortho Solutions LTD, Unit 5 West Station Business Park, Spital Road, Maldon, Essex CM96FF Rx…
Why was Ortho OS Solutions UltOS 3.5mm locking screw 12mm to recalled?
Ortho Solutions Inc is initiating a recall on behalf of UltOS 3.5mm locking screw. The locking thread feature of the screw was not manufactured to specification. As a consequence there is a risk that a screw may not lock onto the plate at the time of primary insertion into the plate and the bone. In the event of this occurring the non functioning screw would need to be replaced with an alternativ
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Part Number (Lot Number) infomration is as follows: OS421712-NS (lot 7891), OS421714-NS (lot 7891) OS421716-NS (lot 7891, 7520), OS421718-NS (lot 7891, 7520) OS421720-NS (lot 7891), OS421722-NS (lot 7891, 7520) OS421724-NS (lot 7891, 7520), OS421726-NS (lot 7891) OS421728-NS (lot 7891), OS421730-NS (lot 7891) OS421735-NS (lot 7520, 7891) ,OS421740-NS (lot 7520, 7891) OS421745-NS (lot 7520, 7891) ,OS421750-NS (lot 7520, 7891) OS421716 (lot 7519) ,OS421718 (lot 7519)
- Product
- Ortho OS Solutions UltOS 3.5mm locking screw 12mm to 50mm Ortho Solutions LTD, Unit 5 West Station Business Park, Spital Road, Maldon, Essex CM96FF Rx…
- Recalling firm
- Ortho Solutions Inc
- Quantity
- 875 units
- Distribution
- US Distribution to states of: CA, CO, GA, IL, LA, NC, NY, TX and WA.
- Recall date
- October 14, 2015
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)