FDAClass IIOctober 15, 2014

Dimension Vista¿ Mass CKMB Isoenzyme Calibrator recall

Dimension Vista¿ Mass CKMB Isoenzyme Calibrator (MMB CAL) an in vitro diagnostic product for the calibration of Creatine Kinase MB Isoenzyme (MMB)…

Why was Dimension Vista¿ Mass CKMB Isoenzyme Calibrator recalled?

Siemens Healthcare Diagnostics has determined that Dimension Vista® Mass CKMB Isoenzyme Calibrator (MMB CAL) shows drift during routine stability monitoring that exceeds our acceptance criteria for drift over the shelf life of this product. Siemens has observed drift from 3% to 12% at MMB concentrations across the assay range. Depending on quality control limits, this drift may not have been d

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

lot 3JD045, exp. 2014-10-01; lot 4BD064, exp. 2015-02-01; lot 4CD012, exp. 2015-03-01; and lot 4DD065, exp. 2015-05-01.

Product
Dimension Vista¿ Mass CKMB Isoenzyme Calibrator (MMB CAL) an in vitro diagnostic product for the calibration of Creatine Kinase MB Isoenzyme (MMB)…
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
6562
Distribution
Worldwided distribution: US (nationwide) and countries of: Australia, Austria, Belgium, Canada, Denmark, Germany, Finland, France, Italy, Japan, Netherlands, New Zealand, Portugal, Saudi Arabia, Slovenia, South Korea, Spain, and Switzerland.
Recall date
October 15, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls