FDAClass IINovember 5, 2014

Proteus 235 Proton Therapy System for cancer treatment recall

Proteus 235 Proton Therapy System for cancer treatment.

Why was Proteus 235 Proton Therapy System for cancer treatment recalled?

While the Proteus 235 is in the pencil beam scanning (PBS) treatment mode, on rare occasions, Low Level Radio Frequency (LLRF) pulsed dee voltage (i.e., voltage applied between two electrodes in the cyclotron to accelerate protons) may generate a hot spot on a pencil beam position within a tumor, generating a dose that might exceed the prescribed dose on a single voxel.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

PAT.107, PAT.109, PAT.112, PAT.113, PAT.114, PAT.115, PAT.116, PAT.119.

Product
Proteus 235 Proton Therapy System for cancer treatment.
Recalling firm
Ion Beam Applications S.A.
Quantity
8
Distribution
Worldwide Distribution-USA (nationwide) including the states of PA, NJ, IL, WA, and TN, and the countries of Germany, Italy, and Czech Republic.
Recall date
November 5, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls